Chlorine Dioxide Analyser System | GF 4632

Chlorine Dioxide Analyser System | GF 4632

Chlorine Dioxide Analyser System – Overview

This integrated all-in-one system measures Chlorine Dioxide residual up to  2 ppm/mg/l. The clear flow cell accommodates the Chlorine Dioxide and optional pH electrode, flow regulator, filter and variable area flow indicator in one compact unit. The flow regulator has a removable filter and accepts inlet pressures of 1 to 8 bar (15 to 120 psi). It maintains constant flow and minimal pressure on the sensors. Water flow is controlled and directed vertically into the sensor tips eliminating the build-up of air bubbles that can cause inaccurate measurement. The flow cell is also designed to maintain a minimum amount of water to ensure sensors stay submerged, even when the system and flow is turned off. This complete system has everything needed to support chlorine monitoring for 1 full year of operation. The design allows quick and easy installation and comes complete with four 4 to 20 mA outputs, flow switch with relay interrupt, four binary inputs and two mechanical relays. Optionally the 9950-X can be used with the 3-950.395-M (159 001 905) Modbus Module.

Chlorine Dioxide Analyser System – Features

  • Reagent free measuring
  • Chemical Guard
  • Prevents overdosing of chemicals
  • Built-in flow switch
  • Chlorine Dioxide and pH electrode performance data (optional pH)
  • Automatic time stamp after successful calibration
  • Customer-enabled alarm feature for re-calibrating
  • Quick and easy installation
  • Built-in flow regulator – constant flow & pressure regardless of inlet pressure

Description

Chlorine Dioxide Analyser System | GF 4632

Specifications
General
Compatible 3-2632-1 Chlorine Dioxide Electrode, 0.02 to 2 ppm / 3-2650-7 Amperometric Electronics
3-2724-00 Flat pH Electrode, 0 to 14 pH / 3-2751-7 pH Sensor Electronics (4632-13 only)
Materials
Panel Black Acrylic
Flow Cell Acrylic
Wiring Enclosure Polycarbonate
Wetted Materials
Flow Cell
Flow Cell, Spacer Rings Acrylic
Flow Regulator Housing Polycarbonate
Strainer, E-clip, Regulator Spring, Float Stainless Steel
Valves, Vent Polypropylene
Flow Cell O-rings, Diaphragm EPR (EPDM), FKM
Chlorine Electrode PVC, PTFE, FKM, Nylon, Silicone
pH Electrode PPS, Glass, UHMWPE, FKM
Flow Switch Polypropylene
Sealing Tape on Valves, Plug and Vent PTFE
Plug Polyethylene
Max. Temperature/Pressure Rating
System Inlet Pressure Rating 1 to 8 bar
Pressure Regulator < 0.69 bar (10 psi) variation over all ranges of flow and pressure
Flow Tolerance ± 15% or rated specification above
Flow Rate Limits 30.24 to 45.36 LPH
Storage Temperature 0 °C to 65 °C
Operating Temperature 0 °C to 45 °C
Electrical
DC Input 24 VDC nominal (12 to 32 VDC, ± 10% regulated)
Environmental
Relative Humidity 0 to 95%
Maximum Altitude 4000 m (13,123 ft)
Enclosure NEMA 4X (with output wire glands sealed)
Shipping Weight 10 kg
Standards and Approvals CE, UL, CUL, FCC, WEEE
China RoHS
Manufactured under ISO 9001, ISO 14001, and ISO 45001