pH/ORP Wet-Tap Electrodes | Signet 2756-2757

pH/ORP Wet-Tap Electrodes | Signet 2756-2757

pH/ORP Wet-Tap Electrodes  – Overview

The Wet-Tap pH and ORP electrodes are sensors which are ideal for a wide range of applications where the installation and removal of the electrode can be performed without the need for system shutdown. The Signet 3719 pH/ORP Wet-Tap Assembly allows the installation and removal of pH or ORP electrodes under process pressure, without process shutdown during routine maintenance and calibration. Process isolation is achieved during electrode retraction with two sets of double O-ring seals on a unique and compact retraction assembly; no separate valve is required. The DryLoc connector with corrosion-resistant gold-plated contacts readily connects the sensor to the mating 2751 pH/ORP Smart Sensor Electronics or the 2760 Preamplifier. The robust polyarylsulphone (PAS) body and choice of bulb pH or flat ORP sensing elements allow a broad range of chemical compatibility for a wide range of applications. The quick temperature response is available in either a Pt1000 or 3 KΩ temperature sensor and allows compatibility with all Signet pH/ORP instruments.

pH/ORP Wet-Tap Electrodes  – Features

  • Installation at any angle – inverted or horizontal
  • PTFE reference junction resists fouling and chemical attack
  • Polyarylsulphone (PAS) body for broad range of chemical compatibility
  • General-purpose bulb pH glass suitable for a wide range of applications
  • Patented DryLoc connector with gold-plated contacts
  • Pt1000 or 3 KOhm Balco temperature element for quick temperature response
  • Electrode removal without process shutdown (when part of Signet 3719 pH/ORP Wet-Tap Assembly)
  • Memory chip-enabled for access to features when connected to the Signet 2751 pH/ORP Smart Sensor

Description

pH/ORP Wet-Tap Electrodes Signet 2756-2757

Specifications
General
Compatibility Signet 3719 Wet-Tap Assembly, 2751 Smart Sensor Electronics or 2760 Preamplifier
Operating Range pH 0 to 14 pH
ORP ±1500 mV
Connector CPVC DryLoc
Temperature Sensor (pH) Pt1000 or 3K Balco for pH
Reference Junctions Porous PTFE
Electrolyte Saturated KCl
Elements Ag/AgCl
Response Time pH < 5s for 95% of signal change
ORP Application dependent
Impedance (pH) < 150 MΩ @ 25 °C
Wetted Materials
Body PAS (Polyarylsulphone)
Reference Junction Porous PTFE
Sensing Surface pH Glass membrane
ORP Platinum
O-rings FKM
Connector CPVC
Max. Temperature Rating
Operating Temperature 0 °C to 85 °C
Recommended Storage Temperature 0 °C to 50 °C
The electrode glass will shatter if shipped or stored at temperature below 0 °C (32 °F)
The performance life of the electrode will shorten if stored at temperatures above 50 °C (122 °F)
Mounting
Any angle is acceptable. Use with 3719 Wet-Tap assembly for mounting electrodes.
Shipping Weight
0.2 kg
Standards and Approvals
Manufactured under ISO 9001 for Quality

Signet Wet-Tap Electrode 3-2756-WTP or 3-2757-WTP with Signet Model 3719 Wet-Tap Assembly

pH/ORP Wet-Tap Electrodes Signet 2756-2757

Brand

Georg Fischer

georg Fischer Signet logo

Reviews

There are no reviews yet.

Be the first to review “pH/ORP Wet-Tap Electrodes | Signet 2756-2757”